www.fdanews.com/articles/183498-medtronic-recalls-insulin-infusion-sets
Medtronic Recalls Insulin Infusion Sets
September 13, 2017
Medtronic recalled some of the infusion sets used with its insulin pumps due to problems with a discontinued component.
The company said the problem involves a vent membrane that’s susceptible to blockage during priming and fill-tubing.
No pumps or glucose sensors are affected by the recall. Infusion sets made since April 2017 use an updated vent membrane design.