Home » FDA Provides Update on Dräger’s Recall of Its Breathing System Filter

FDA Provides Update on Dräger’s Recall of Its Breathing System Filter

June 15, 2022

The FDA has issued an update on Dräger’s May 16 recall of its SafeStar 55 breathing system filter with lot number LT2103, deeming it a Class 1 recall because of the risk of serious injury or death.

The reason for the recall is that an obstructed SafeStar 55 breathing system filter was used on a patient during anesthesia who “reportedly became hypoxic and had to be reanimated,” according to the company.

The manufacturing problem was linked to “a human error” while equipment was being serviced. An employee mistakenly placed a filter that had failed a resistance test in the wrong box, the company explained, adding that most likely only one filter was affected, but it is recalling the entire lot to be safe.

View today's stories

Medical Devices Postmarket Safety

Upcoming Events

  • 13Jul

    FDA Ad and Promo Enforcement Trends: Past, Present and Future

  • 28Jul

    EU-MDR’s Growing Pains: Are You Current with the Latest Developments?

  • 28Jul

    Is Your Product a Drug or Device — and Does It Matter?: A Recent Appeals Court Decision Says It Does

  • 02Aug

    Medical Device Cybersecurity: Understand the Latest Developments

  • 03Aug

    Pharmaceutical Process Validation: Best Practices for Success

  • 10Aug

    Gene & Cell Therapy Regulation: Comparability and Other New Developments

Featured Products

  • FDA’s New Quality System Regulation: Transitioning from QSR to ISO 13485

    FDA’s New Quality System Regulation: Transitioning from QSR to ISO 13485

  • Selecting and Implementing Electronic Document Management Systems in the EU

    Selecting and Implementing Electronic Document Management Systems in the EU

Featured Stories

  • FDA Provides Update on GE Healthcare’s Recall of Its Carescape R860 Ventilators

  • Gilead Resubmits NDA for Lenacapavir Following Complete Response Letter

  • FDA Warns Baby Neck Floats Used in Water Therapy Can Cause Death or Injury

  • U.S. Government Inks $3.2 Billion Contract With Pfizer-BioNTech for Vaccine Booster Campaign

The Revised ICH E8: A Guide to New Clinical Trial Requirements

Learn More
  • Drug Products
    • Quality
    • Regulatory Affairs
    • GMPs
    • Inspections and Audits
    • Postmarket Safety
    • Submissions and Approvals
    • Research and Development
    • Commercial Operations
  • Device Products
    • Quality
    • Regulatory Affairs
    • QSR
    • Inspections and Audits
    • Postmarket Safety
    • Submissions and Approvals
    • Research and Development
    • Commercial Operations
  • Clinical Products
    • Trial Design
    • Data Integrity
    • GCP
    • Inspections and Audits
    • Transparency
  • Privacy Policy
  • Do Not Sell My Personal Information
Footer Logo

300 N. Washington St., Suite 200, Falls Church, VA 22046, USA

Phone 703.538.7600 – Toll free 888.838.5578

Copyright © 2022. All Rights Reserved. Design, CMS, Hosting & Web Development :: ePublishing