www.fdanews.com/articles/122899-medtronic-cardiac-unit-warned-for-poor-purchasing-controls
Medtronic Cardiac Unit Warned for Poor Purchasing Controls
December 11, 2009
Medtronic’s cardiac rhythm management
unit has received a warning letter for
purchasing control failures and other GMP deficiencies.
FDA inspectors were not able to find any record that indicated a performance review was done on a supplier between 2002 and 2007 even though Medtronic’s supplier quality management strategy required regular evaluations, according to the Nov. 9 letter posted recently to the agency website.
Further, the Mounds View, Minn., facility only tested one lot of a particular component to qualify it, which violated the company’s
component/material qualification and characterization procedure.
The GMP Letter