King Systems Corp. - September 4, 2009
Issue Date
Sep 04, 2009
Region
Central
Category
Device
$117.00
Address 1
15011 Herriman Blvd
City
Noblesville
State
IN
Zip/Postal Code
46060
Country
US
Inspection Range
Sep 01, 2009 - Sep 04, 2009
Pages
2
First Observation
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Investigator(s): Calvin K. Cook