Spray-Tek, Inc. - June 3, 2010
Issue Date
Jun 03, 2010
Region
Central
Category
Drug
$117.00
Address 1
3010 Avenue B
City
Bethlehem
State
PA
Zip/Postal Code
18017
Country
US
Inspection Range
May 20, 2010 - Jun 03, 2010
Pages
1
First Observation
Laboratory methods used for particle size analysis and moisture determination have not been validated. In addition, there is no evidence the following analytical equipment was qualified: High Performance Moisture Analyzer, serial number and the serial number.
Investigator(s): Edward D. McDonald